Safety Alert: Class II: Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
healthClass IIRecalled on Sep 17, 2025
Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.
🔢 192
Sold At: WASATCH PHOTONICS(U.S.)
Recall #: FDA-DEVICE-Z-2478-2025
Source: FDA-ENFORCEMENT
Product Identification
- Lot/Batch Codes:
- all codes included in the model numbers.
Get AI-Powered Summary
Premium members get instant insights to help you understand recalls faster
Instant 1-3 sentence summary
Quick action bullets
Clear guidance on what to do
Summary
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
What to do next
- 1Stop using this product - may cause temporary or medically reversible health problems
- 2Check with retailer or manufacturer for recall status and remedy options
- 3Contact WASATCH PHOTONICS for return or refund information
- 4Visit the official FDA recall page for complete details