SureStep Foley Tray System BARD¿ Lubri-Sil Foley Catheter Tray, REF: A942216
Sold At: C.R. Bard Inc(US Nationwide distribution including in the states of NC, FL, NY, MA, MN, TX, CA, OH, MI, WA, NJ, IL, ID, SC, IN, OK, OR, AL, VA, HI, MO, MS, MD, AZ, KY, GA, IA, CO, NM, PA, ME, UT, NE.)
Recall #: FDA-DEVICE-Z-2591-2025
Source: FDA-ENFORCEMENT
Product Identification
- Lot/Batch Codes:
- Lot# NGJX4320
Get AI-Powered Summary
Premium members get instant insights to help you understand recalls faster
Summary
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction
What to do next
- 1Stop using this product - may cause temporary or medically reversible health problems
- 2Check with retailer or manufacturer for recall status and remedy options
- 3Contact C.R. Bard Inc for return or refund information
- 4Visit the official FDA recall page for complete details
Summaries only; always follow the official notice.