healthClass IIRecalled on Oct 1, 2025

Philips CT systems labeled as the following: 1. IQon Spectral CT, Model Number: 728332.

🔢 81 units

Sold At: Philips North America Llc(Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.)

Recall #: FDA-DEVICE-Z-2601-2025

Source: FDA-ENFORCEMENT

Philips CT systems labeled as the following:    1. IQon Spectral CT, Model Number: 728332.

Product Identification

Lot/Batch Codes:
1. Model Number: 728332; UDI-DI: 00884838059542; Serial Numbers: 860088, 860023, 860077, 860086, 60019, 860152, 860215, 60031, 60123, 60129, 60152, 860011, 860049, 860019, 860008, 860030, 860025, 860009, 860022, 860013, 860024, 860020, 860036, 860041, 860046, 860055, 860058, 860054, 860073, 860065, 860070, 860072, 860066, 860103, 860076, 860085, 860095, 860102, 860101, 860104, 860175, 860136, 860117, 860120, 860121, 860124, 860125, 860191, 860131, 860150, 860164, 60041, 860207, 60003, 860157, 860167, 860158, 860196, 60012, 860176, 860185, 860168, 860179, 860194, 60010, 60013, 60007, 60001, 395003, 860209, 860208, 60036, 60030, 60027, 60054, 60055, 60053, 60058, 60079, 395005, 60145.

Get AI-Powered Summary

Premium members get instant insights to help you understand recalls faster

Instant 1-3 sentence summary
Quick action bullets
Clear guidance on what to do
Learn About Premium

Summary

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

What to do next

  • 1Stop using this product - may cause temporary or medically reversible health problems
  • 2Check with retailer or manufacturer for recall status and remedy options
  • 3Contact Philips North America Llc for return or refund information
  • 4Visit the official FDA recall page for complete details

Summaries only; always follow the official notice.