healthClass IIRecalled on Sep 24, 2025

Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.

🔢 2,908 units

Sold At: Ortho-Clinical Diagnostics, Inc.(US Nationwide distribution.)

Recall #: FDA-DEVICE-Z-2559-2025

Source: FDA-ENFORCEMENT

Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.

Product Identification

Lot/Batch Codes:
Catalog number 8433880; UDI# 110758750004911; Coating 0130 and above, this includes Lot numbers 3533-0130-2317, 3533-0130-2717, 3533-0130-2718, 3533-0130-2719, 3533-0130-4546, 3533-0130-4605, 3533-0131-4763, 3533-0131-6114, 3533-0131-6115, 3533-0131-6116, 3533-0131-6128; Expiration date range 01-Sep-2026 to 01-Jan-2027

Get AI-Powered Summary

Premium members get instant insights to help you understand recalls faster

Instant 1-3 sentence summary
Quick action bullets
Clear guidance on what to do
Learn About Premium

Summary

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

What to do next

  • 1Stop using this product - may cause temporary or medically reversible health problems
  • 2Check with retailer or manufacturer for recall status and remedy options
  • 3Contact Ortho-Clinical Diagnostics, Inc. for return or refund information
  • 4Visit the official FDA recall page for complete details

Summaries only; always follow the official notice.